MRI Space Planning
1. MRI Rooms Space Planning
Technical
Finished-Room Size
|
Technical
Finished-Room High
|
|
Control room
|
min. 150 x 190 cm
|
min. 210 cm
|
Equipment room
|
min. 230 x 165 cm
|
min. 220 cm
|
Examination Room
• Full Body
Mobile patient table
|
min. 351 x 662 cm
|
min. 240 cm
|
2. Area requirements and floor loading for the magnet
The system has to be installed on a solid
underground with sufficient carrying capacity,
such as, e.g. concrete. The load bearing
capacity has to be checked by a stress analyst.
The floor in the vicinity of the
magnet must be levelled to within max. ± 2 mm.
External vibration or shocks
affecting the magnet may degrade image quality. In the 3 spatial orientations
the building vibration must not exceed the following specification:
2.1 Building vibration specification:
a max = - 80 dB(g) in the
frequency range from 0 to 100Hz.
The requirement for a max is
frequency dependent.3. Weights to be consider
Magnet:
|
As
per Manufacturer,~ 5500 kg
|
Patient
table:
|
~250kg
|
Electronic
cabinets
|
~1850 kg
|
UPS
to support the MRI system
|
~2500 kg
|
Iron
Shielding
|
As
per static calculation
|
RF
Cabin
|
Depends
on the material and size
|
4. MRI Environmental Requirements
MRI
|
Examination
room
|
Equipment
room
|
Control room/
Evaluation
|
Room temperature
Temperature gradient
|
18 to 22 °C
n.a.
|
15 to 30 °C
</= 1.0 K/5min
|
15 to 30 °C
n.a.
|
Relative humidity
|
40 to 60 %
|
40 to 80 %
|
40 to 60 %
|
Absolute humidity
|
< 11 g / kg
|
< 11 g / kg
|
< 11 g / kg
|
5. Room lighting in the MRI room
The
magnetic field adversely affects the operating life of light bulb located in
the
immediate
vicinity of the magnet. The filament in the light bulb oscillates with the frequency
of the power supply.
It is therefore
recommended to connect light fixtures in the vicinity of the magnet to
a DC
voltage supply. If room lighting is supplied with DC voltage, correct polarity
ofthe sockets should be ensured during their installation.
Residual AC ripple should be £ 5 %.
Fluorescent
lighting must not be installed inside the examination room. Only lamps without
phase angle control should be used. Do not use energy-saving lamps.
6. Noise
Emission
If required, noise reduction should be realized based on the noise emission
values as specified.
MRI
|
Examination room
|
Control room
|
Equipment room
|
Average values across 8 hours
|
≤ 80 dB(A)
|
≤ 54 dB(A)
|
≤ 67 dB(A)
|
7. RF Shielding
An RF
shielding (Faraday cage) is required for the MR examination room. This
shielding
protects
the environment from RF interference and conversely protects the MR system
from
external interference.
Required
attenuation: >90 (Co-siting 100) dB over the frequency range 15 to 128 MHz.
These
values must be verified by measurement before the MR system is installed.
RF
shielding components (doors, windows, interfaces) and complete modular RF
cabins
can be
supplied from different sources.
8. Quench Pipe
A quench pipe (thermally insulated tube)
made of non-magnetic metal must be fitted
from the
super-conducting magnet to the outside of the building in order to vent thevaporising helium gas.
9. Magnetic field level warning signs in the control zone ³ 0.5 mT
If the magnetic flux density in a given area
exceeds 0.5 mT, it is necessary to display
warning signs and restrict access in accordance
with local regulations.
10. Water connection
Either you
connect a chiller recommended by the MRI manufacturer or you can connect it to
hospital chiller system.
11. Power Supply:
The Superconductive magnet normally required 3Phase/N/PE, AC 380V, Connections Value ~120 KVA.
12. Guidelines for max. permissible flux density (mT)
Max.
permissible flux density
|
mT
|
|
Servoventilator,
Defibrillator
|
20
|
|
Filter
|
10
|
|
Small motors, watches,
photographic equipment,
magnetic data storage
devices (short term)
|
3
|
|
Computers, disc drivers,
oscilloscopes, host PC, imager PC
|
1
|
|
B/W monitors, X-ray tubes,
storage of magnetic data carriers,
cardiac pacemakers, insulin
pumps.
(safety limit for
unrestricted open access)
|
0.5
|
|
CT-systems
|
0.2
|
|
Colour monitors (CRT)
|
0.15
|
|
Linear accelerators
|
0.1
|
|
Image intensifiers, gamma cameras linear
accelerators
|
0.05
|
|
Laser cameras /
digital cameras
|
see manual of the manufacturer
|
|
be reduced by a magnetic shielding. Typical lines of constant magnetic flux density are
shown in the drawing.
No comments:
Post a Comment